Esterline Pressure Systems Logo header spacer Pressure Systems Site Map Pressure Systems Home
 
 

ESP: Heater Chamber Temperature Adjustment Procedure

  return to ESP Scanners main product page

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
 


The Resistance Temperature chart below gives the characteristic of the temperature sensor used in the ESP Heater Chamber.

The following procedure can be used to adjust the heater to a different temperature (less than 80ºC scanner operating temperature).

  1. Determine a control set point temperature fitting the application.
  2. Connect an Ohmmeter to the two test points on the Heater Chamber.
  3. Adjust the potentiometer until the value read from the chart is indicated on the meter.
  4. Note 1: This procedure results in a temperature setting within ±3º of the desired setpoint.

    Note 2: The heating element in the Heater Chamber can develop about 33 watts of power. The chamber provides insulation, however there are limitations for achievable regulation. The Heater Chamber reduces the temperature effect to the scanner by a factor of 10:1.

    Note 3: It is recommended to set the Heater Chamber temperature setpoint to 5ºC above the expected ambient temperature. Maximum scanner temperature environment is 80ºC.

    Note 4: For operation in a very cold environment (-50ºC), it is important to set the temperature to above freezing (0ºC), so water condensation does not occur.

    Note 5: Factory default temperature setpoint is 40ºC.

DEG C

CONTROL SETPOINT IN OHMS

DEG C

CONTROL SETPOINT IN OHMS

0

43.8225

42

54.09291

3

44.53217

45

54.79913

6

45.24462

48

55.55962

9

45.95994

51

56.32418

12

46.67822

54

57.09291

15

47.39955

57

57.86588

18

48.124

60

58.64319

21

48.85167

63

59.42491

24

49.58264

66

60.21115

27

50.317

69

61.00197

30

51.05483

72

61.79748

33

51.79623

75

62.59774

36

52.54127

78

63.40287

39

53.29005

81

64.21291

 

No. 0704

 
Pressure Systems
34 Research Drive
Hampton, VA 23666
800.678.7226
KPSI Selection Guide Quality Policy
ISO-9001:2000
Receive E-Newsletter
IPC Member Logo WEEE/RoHS Statement
WEEE/RoHS
Statement
 


Home | Environmental Products | Industrial Products | Aerospace Products | Buy Now
KPSI Level and Pressure Selection Guide | Quartzonix Pressure Standard | NetScanner System | System 8400 |
Electronic Pressure Scanners | S8256 System | Miscellaneous Hardware | News Archive
Contact Us | Distributors | Links | Employment | Register | Search