| Parameter |
9816 |
98RK-1
|
Units |
Comments |
PNEUMATICS |
| Pressure Ranges (1) |
10 (2.5)
20 (5.0) |
N/A |
inches WC (kPa) |
ranges < 100 psi (689 kPa) are bidirectional
ranges >= 100 psi (689 kPa) are unidirectional |
|
1 (7.0)
2.5 (17)
5 (34)
10 (69)
15 (103)
30 (207)
45 (310)
100 (689)
250 (1724)
500 (3447)
750 (5171)
850 (5860) |
N/A |
psid (kPa) |
Number of Measurement Inputs |
16 |
N/A |

|
 |
| Pneumatic Fittings |
.063 bulged tubulation, or
SAE 5/16-24 o-ring boss (if front access specified) |
SAE 7/16-24 o-ring boss |

|
variety of compression adapter fittings available
See page 6 for standard fittings |
| Media |
Dry, non-corrosive gas |

|
for liquid pressures refer to
Model 9022 datasheet |
| Proof Pressure (2) |
3.0x 2.0X
1.5X |
N/A |
FS min |
range < 15 psi (103 kPa)
range 15-100 psi (103-689 kPa)
range > 100 psi (689 kPa) |
| Burst Pressure |
5.0X
3.0X
2.0X |
N/A |
FS min |
range < 15 psi (103 kPa)
range 15-100 psi (103-689 kPa)
range > 100 psi (689 kPa) |
| Max Line Pressure (3) |
100 (689)
600 (4134) |
N/A |
psi (kPa)
psi (kPa) |
range <= 5 psi (34 kPa)
range > 5 psi (34 kPa) |
Manifold Control
Pressure |
80 (552)
125 (862) |
psi (kPa) |
min max |
| Purge Pressure (4) |
greater max input pressure
850 (5860) |
psi (kPa) psi (kPa) |
min max |
STATIC PERFORMANCE |
| Measurement Resolution |
±0.003 |
N/A |
% FS |

|
Static Accuracy (5) After Rezero |
±0.05
±0.15 |
N/A |
% FS
% FS |
range > 2.5 psi (17 kPa)
range <= 2.5 psi (17 kPa) |
| Total Thermal Error (7) |
±0.001 ±0.0015 |
N/A |
%FS/șC %FS/șC |
range > 2.5 psi (17 kPa)
range <= 2.5 psi (17 kPa) |
| Line Pressure Effect |
±0.0003
±0.0008
±0.001 |
N/A |
% FS/psi
%
FS/psi
% FS/psi |
range <= 100 psi (689 kPa)
range >5 psi (45 kPa) <100 psi (689 kPa)
range <= 5 psi (34 kPa) |
| Measurement Rate |
100 |
meas/ch/sec |
engineering units |
COMMUNICATION |
| Ethernet |
10BaseT
half duplex
|
10/100/1000 BaseT
auto negotiating |

|
|
| Protocol |
TCP
and UDP |
% FS
|
static or dynamic IP addressing |
ELECTRICAL |
| Input Voltage Options |
N/A
|
85-264 |
VAC |
47-63 Hz |
| Operating Power |
N/A |
65 (7)
|
watts |
 |
| Hardware Trigger |
2.5
|
VDC |
TTL compatible
differential input, ±5 VDC common mode voltage |
ENVIRONMENTAL/PHYSICAL |
| Calibrated Temp Range (10) |
0 to
60 |
șC |
|
| Operating Temp Range |
0 to 50 |
șC |
up to 95% non-condensing humidity |
| Storage Temp |
-20 to 80 |
șC |
 |
| Size |
2.06 x 7.00 x 15.25
(5.2 x 17.7 x 38.7)
|
17.00 x 8.75 x 19.00
(43.2 x 22.2 x 48.3)
|
inches (cm) |
L x W x H |
| Weight |
6.5
(3.0) |
26.9
(12.2)
|
lb (kg) |
 |