Esterline Pressure Systems Logo header spacer Pressure Systems Site Map Pressure Systems Home
 
 

Quartzonix™ Series 960 Pressure Standard: Specifications

link arrow  return to Series 960 main page

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
   
Parameter
960
Units
Comments

PNEUMATICS

Pressure Ranges
10 - 16 (69 - 110)
15 (103)
23 (159)
30 (207)
45 (310)
65 (448)
100 (689)
200 (1379)
300 (2068)
500 (3447)
psia (kPa)
Barometric only
Proof Pressure (1)
1.5
x FS
Burst Pressure (2)
2.0
x FS
Pressure Media
Media compatible with 316SS, nickel and solder

Pressure Fitting
1/8" compression

STATIC PERFORMANCE

Resolution (3)
±0.0001
% FS
Pressure Hysteresis
±0.005
% FS
Static Accuracy (4)
±0.01
% FS
Maximum Deviation
over Temperature (5)
±0.015
% FS
over specified temperature range
Thermal Hysteresis
±0.005
% FS
over specified temperature range
Long Term Drift
±0.01
% FS 
6 months, maximum

ELECTRICAL

Pressure Signal (6)
Temperature Signal
45-30
172
kHz
kHz
nominal
+45 ppm/șC
Output Signal
5
volt
P-P
square wave
capacitively coupled
Turn on Time
100
m sec
typical
Power Requirements
8-15
VDC
1 mA
Electrical Connection
8 ea. #30 AWG flying leads

ENVIRONMENTAL

Calibrated Temp Range
0 to 60
ºC  
Storage Temp -20 to 70 ºC  
Acceleration Sensitivity
±0.004
% FS/g worst axis
Vibration Sensitivity
±0.001
% FS/g
20g peak, 10 Hz - 2 KHz
Maximum Shock
200
g/10
msec
1/2 sine

PHYSICAL

Size
1.73 x 1.73 x 1.35
(4.4 x 4.4 x 3.4)
in
(cm)
Weight
9.5 (270)
oz. (gms)
(1) Maximum applied pressure without causing a calibration shift.
(2) Maximum applied pressure without causing permanent damage to quartz sensing element.
(3) Achievable resolution using gate time of 1/2 second for frequency counter with 20,000 MHz time base. Increased or decreased resolution may be obtained be lengthening or shortening gate times.
(4) Calibration conformance to a secondary pressure standard after calibration.
(5) Maximum error at any pressure within calibrated temperature range using supplied calibration coefficients/equations.
(6) Nominal 10% frequency change with full scale pressure.

Specifications subject to change without notice.
 

 
Pressure Systems
34 Research Drive
Hampton, VA 23666
800.678.7226
KPSI Selection Guide Quality Policy
ISO-9001:2000
Receive E-Newsletter
IPC Member Logo WEEE/RoHS Statement
WEEE/RoHS
Statement
 


Home | Environmental Products | Industrial Products | Aerospace Products | Buy Now
KPSI Level and Pressure Selection Guide | Quartzonix Pressure Standard | NetScanner System | System 8400 |
Electronic Pressure Scanners | S8256 System | Miscellaneous Hardware | News Archive
Contact Us | Distributors | Links | Employment | Register | Search