Esterline Pressure Systems Logo header spacer Pressure Systems Site Map Pressure Systems Home
 
 

NetScanner™ System: 9032/9034/9033/9038 Specifications

link arrow  return to 9032/9034/9033/9038 main page

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
 
Parameter
9032/33
9034/38
Units
Comments

PNEUMATICS

Reference Pressure Format
Absolute
Differential
   
Pressure Range
Absolute
(Model 9032 and 9034)

11-16 (76 -110)
15 (103)
23 (159)
30 (207)
45 (310)
65 (448)
100 (689)
200 (1379)
300 (2068)
500 (3447)
750 (5171)

±1 (7)
±2 (14)
±5 (21)

psi (kPa)

 
Media
Dry, non-corrosive gas

Output Range (1)
N/A
5 to 100
N/A
-100 to 100
% FS
9038, see reference pressure
Source Gas(2)
 
 
 
Min
N/A
FS + 10 (69)

1.1
N/A
FS + 10 (69)

1.1

psi (kPa)

x FS

range <=100 psi (689 kPa)

range > 100 psi (689 kPa)
Max
N/A
250 (1724)

1.5
N/A
250 (1724)

1.5

psi (kPa)

x FS

range <=100 psi (689 kPa)

range > 100 psi (689 kPa)
Vacuum(3)
N/A
2 (60)
N/A
2 (60)
scfm
lpm
 
Output Flow Capacity
N/A
0.1 (3)
N/A
0.1 (3)
scfm (lpm)
min
Overpressure (4)
(input/output/ref ports)
1.3
x FS
max

STATIC PERFORMANCE

Measurement Resolution
±0.0005
±0.003
%FS
 
Relative Accuracy (5)
±0.01
±0.025
%FS
 
Long Term Stability ±0.01 ±0.05 (6) % FS/6 mos  
Thermal Stability (7) ±0.001 % FS/șC from 25șC
Measurement Rate (8) 10 readings/sec  
Controller Setpoint Resolution N/A 0.003 N/A 0.003 % FS  
Output Setting Time (9) N/A 5 N/A 5 sec ±0.01% of final value

DIGITAL INPUTS/OUTPUTS

Input Voltage 18 - 36 VDC unregulated
Input Current 250
850
mA @ 24 VDC, quiescent
@24 VDC, active
Hardware Trigger Threshold 2.5 VDC TTL compatible differential input, ±5 VDC common mode voltage
Digital Input/Output
  Output Voltage (10)

  Sink Current per Output

  Max Source Current (11)

18-36

0.5

1.0

VDC

A

A

max

max

max @ +24 VDC

COMMUNICATION

Communication 10 BaseT Ethernet % FS  
Protocol TCP and UDP % FS static or dynamic IP assignment

ENVIRONMENTAL/[PHYSICAL

Calibrated Temp Range
0 to 50
șC
see factory for extended temperature range
15 to 45
ºC
range = 5 psid
Operating Temp Range
-10 to 60
șC
up to 95% non-condensing humidity
0 to 50
ºC range = 5 psid
Storage Temp
-20 to 80
șC  
0 to 70
ºC range = 5 psid
Size
9.5 x 3.5 x 3.875
(24.1 x 8.9 x 9.5)

9.5 x 6.375 x 3.85
(24.1 x 16.2 x 9.8)
inches
(cm)

L x W x H, range < 750 psi


L x W x H, range >= 750 psi
range = 5 psid

Weight 4.0 (1.8)
5.8 (2.3)
N/A
5.2 (2.3)
7.0 (3.2)
N/A
4.0 (1.8)
N/A
7.4 (3.6)
5.2 (2.3)
N/A
8.6 (3.9)

lb (kg)
lb (kg)
lb (kg)

range < 750 psi (5171 kPa)
range >= 750 psi (5171 kPa)
range = 5 psid (21 kPa)
(1) Control of subatmospheric pressures requires vacuum pump.
(2) Gas should be pre-filtered to 10 microns, free of oil and have dew point of -25șC.
(3) Required for control of subatmospheric pressure, recommended for control near zero psig (kPa).
(4)

For absolute reference units (9032 & 9034), overpressure is defined as the maximum absolute pressure which may be applied to the unit without damage.  For differential units (9033 & 9038), overpressure is defined as the maximum pressure differential between the pressure standard input and reference.

(5) Relative accuracy incldues the combined errors due to non linearity, hysteresis and non-repeatabililty.
(7) After internal rezero.
(8) Includes effects of zero and span.
(9) Measurement rate to host computer depends on host computer software throughput rate. Consult users manual for more information.
(10) Setting rate depends on pneumatic volume of tubing and unit under test connected to instrument.
(11) Output voltage directly equivalent to 903x input voltage supply.
(12) Maximum sink/source current for all external control outputs.

 

Specifications subject to change without notice.

 

 
 
Pressure Systems
34 Research Drive
Hampton, VA 23666
800.678.7226
KPSI Selection Guide Quality Policy
ISO-9001:2000
Receive E-Newsletter
IPC Member Logo WEEE/RoHS Statement
WEEE/RoHS
Statement
 


Home | Environmental Products | Industrial Products | Aerospace Products | Buy Now
KPSI Level and Pressure Selection Guide | Quartzonix Pressure Standard | NetScanner System | System 8400 |
Electronic Pressure Scanners | S8256 System | Miscellaneous Hardware | News Archive