Esterline Pressure Systems Logo header spacer Pressure Systems Site Map Pressure Systems Home
 
 

NetScanner™ System: 9016 Specifications

link arrow  return to 9016 main page

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
 
Parameter
9016

Units
Comments

PNEUMATICS

Pressure Range (1)
10 (2.5)
20 (5.0)
In WC (kPa)

ranges greater than or equal to 100 psi (689 kPa) are unidirectional only

1 (7.0)
2.5 ( 17)
5 (34)
10 (69)
15 ( 103)
30 (207)
45 (310)
100 (689)
250 (1724)
500 (3447)
750 (5171)
850 (5860)
psid / kPa
Number of Pressure
Inputs
16

Pneumatic Fittings
1/16", 1/8", 1/4"
5/16" - 24 SAE
.063", .040" bulge tube
compression fittings
o-ring boss
QDC tubing plate
Media
Dry, non-corrosive gas

Proof Pressure (2)
3.0X
2.0X
1.5X
FS min
range less than 15 psi (103 kPa)
range 15-100 psi (103-689 kPa)
range greater than 100 psi (689 kPa)
Burst Pressure
5.0X
3.0X
2.0X
FS min
range less than 15 psi (103 kPa)
range 15-100 psi (103-689 kPa)
range greater than 100 psi (689 kPa)
Max Line Pressure (3)
100 (689)
600 (4137)
psi (kPa)
psi (kPa)
range less than or equal to 5 psi (34 kPa)
range greater than 5 psi (34 kPa)
Manifold Control
Pressure
80 (552)
125 (862)
psi (kPa)
min
max
Purge Pressure (4)
greater than max input pressure
850 (5860)
psi (kPa)
min
max
STATIC PERFORMANCE
Measurement
Resolution
±0.003
% FS
Static Accuracy (5)
After Rezero
±0.05
±0.15
% FS
% FS
range greater than 2.5 psi (17 kPa)
range less than or equal to 2.5 psi (17 kPa)
Total Thermal Error (6)
±0.001
±0.0015
%FS/șC
%FS/șC
range greater than 2.5 psi (17 kPa)
range less than or equal to 2.5 psi (17 kPa)
Line Pressure Effect
±0.0003
±0.0008

±0.001
% FS /psi
% FS/psi

% FS/psi
range greater or equal to 100 psi (689 kPa)
range greater than 5 psi (34 kPa) to
   range less than 100 psi (689 kPa)
range less than or equal to 5 psi (34 kPa)
Measurement Rate
100
meas/ch/sec
engineering units

COMMUNICATION

Ethernet
10BaseT

half duplex
Protocol
TCP and UDP

static or dynamic IP assignment

ELECTRICAL

Input Voltage
18 - 36
VDC unregulated
Input Current
200
mA @ 24 VDC
Hardware Trigger
Threshold
2.5
VDC TTL compatible differential
input, ±5 VDC common
mode voltage

ENVIRONMENTAL/PHYSICAL

Calibrated Temp
Range (8)
0 to 50
șC
consult factory for extended temps
Operating
Temp Range
-20 to 70
șC
up to 95% non-condensing humidity
Storage Temp
-20 to 80
șC
Size
9.50 x 3.50 x 3.537
(24.1 x 8.9 x 8.98)
inches
(cm)
L x W x H
Weight
6.5
(2.95)
lb
(kg)
Splashproof Rating

NEMA 13

   
Acoustic Noise Rating

160dB

   
Shock Rating

30

g

 
(1) Standard ranges listed, consult factory for special ranges.
(2) Maximum applied pressure without reducing sensor performance
(3) For anticipated Line (Reference) Pressures . . .
  • above 100 psi (689 kPa) on FS Pressure Ranges of less than or equal to 5 psi (34 kPa) or
  • above 600 psi (4137 kPa) on FS Pressure Ranges of greater than 5 psi (34 kPa)
  • calibration manifold should not be shifted at line pressure
  • rate of line pressure change should be no more than 10 psi/sec
(4) Purge pressure must be greater than maximum input pressure
(5) Static accuracy includes the combined errors due to non-linearity, hysteresis and non-repeatability
(6) Includes thermal effects of zero & span relative to 25șC
(8) Consult factory for special temperature range calibration
 

Specifications subject to change without notice.

 

 
 
Pressure Systems
34 Research Drive
Hampton, VA 23666
800.678.7226
KPSI Selection Guide Quality Policy
ISO-9001:2000
Receive E-Newsletter
IPC Member Logo WEEE/RoHS Statement
WEEE/RoHS
Statement
 


Home | Environmental Products | Industrial Products | Aerospace Products | Buy Now
KPSI Level and Pressure Selection Guide | Quartzonix Pressure Standard | NetScanner System | System 8400 |
Electronic Pressure Scanners | S8256 System | Miscellaneous Hardware | News Archive
Contact Us | Distributors | Links | Employment | Register | Search