System 8400 Updates: FIU V2.25 Fiber-Optic Diagnostic Test

link arrow  return to System 8400 Updates

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

1.0 Introduction

This document describes two new diagnostic test commands that have been added to the latest version (V2.25) of the FIU module code. These diagnostic tests allow customer field testing of the FIU/SDI fiber-optic link.

2.0 Description

The previous version of the FIU module code (V2.24) does not perform any diagnostic tests of the fiber-optic link; it also does not check for link errors at any time during system operation. Two new diagnostic tests have been added:

  1. The "loop-back" test, where the FIU repeatedly sends a varying pattern of bits to the SDI, and with the SDI configured to 'echo' back whatever it receives, the FIU compares 'sent' data with 'received ' data.
  2. The 'SDI' test, where the FIU commands the SDI to read a value that isn't expected to vary by more than a small ( and predictable) amount; data is then read repeatedly from the SDI and checked to see if it falls within the expected range.

The existing version of the SDI module doesn't provide ability to 'echo' data on command. To perform the 'loop-back' test, the cable must be changed to either one of these configurations.

  1. A short length of cable connected between the 'data out' and data in' connectors on the front of the FIU module. This tests only the FIU.
  2. A short length of cable connected between the two ends of the normal FIU fiber-optic (in place of the SDI module). This tests the FIU and the installed cable. NOTE: If a short length of cable to couple the ends is not available, each leg of the fiber cable can be tested by connecting one end to the 'data in' and the other end to 'data out' on the FIU. This will require the test be run twice and can only be done when both ends of the fiber-optic cable are at the FIU location.

To perform the 'SDI' test, the normal FIU-to-SDI cables are used, and any one of the following test methods/setups can be selected:

  1. DTC excitation voltage test (one DTC scanner needed)
  2. Analog Ground reference test (no scanner needed)
  3. M0 logic zero test (no scanner needed)
  4. M1 logic one test (no scanner needed)

The 'loop-back' test and 'SDI' tests are both selected using the XC1 host command. Each test is executed repeatedly a user-selected number of times, during which any fiber-optic link errors that occur are counted and received data is checked for validity. On completion, the results are returned to the host.

3.0 User Interface

The 'loop-back' test is selected using the following command:

XC1 crs 6 13 passcount timeout
where:
crs = FIU module's cluster, rack and slot address
passcount = number of times test is repeated (1-65535)
timeout = maximum delay waiting for a reply ( valid range 1-100 microseconds; 20 recommended)

On completion, the following information is returned in the command's response (part binary, part ASCII string):

xxxxxxxx "dataerrors timeouts mintime maxtime"
where:
xxxxxxxx = 8 bytes of binary header packet information (See section 5.12 of 8400 User's Manual) This binary information can generally be ignored.
dataerrors = number of times 'sent' data did not match the 'received' data
timeouts = number of timeout errors (no data was received)
mintime = minimum reply time seen (will usually be 0 or 1 if no timeout, > if error)
maxtime =maximum reply time seen (will usually be 0 or 1 if no timeout, > if error)

For example, if the FIU module is in Slot 1 and this command is entered:

XC1 111 6 13 50000 20

The returned response might be:

xxxxxxxx "0 0 0 1"
which would indicated the fiber-optic cable is operating properly.

If the returned response is:

xxxxxxxx "12345 42 0 13"

then the non-zero error counts and/or timeouts indicate an intermittent hardware problem, either in the cable or in the FIU module.

The 'SDI' test is selected using the following command:

XC1 crs 6 14 passcount timeout scannertype delaytime minvalue maxvalue
where:
crs = FIU module's cluster, rack and slot address
passcount = number of times test is repeated (1-65535)
timeout = maximum delay (range is 1-100 microseconds; 20 recommended)

scanner type is one of the following:
1 to 16 = performs the 'DTC excitation voltage' test using the specified scanner #
0 = performs the 'Analog Ground Reference' test
-1 = performs the 'M0 Logic Zero' test
-2 = performs the 'M0 Logic One' test
delaytime = minimum delay in microseconds between commands sent to the SDI (range is 1-100; 20 recommended)
minvalue = minimum data value (0-65535 counts)
maxvalue = maximum data value (0-65535 counts)

On completion, the following information is returned in the command's response (part binary, part ASCII string):

xxxxxxxx "timeouts mintime maxtime minerrors maxerrors average min max
where:
xxxxxxxx = 8 bytes of miscellaneous binary packet header information (ignore)
timeouts = number of timeout errors (no data was received)
mintime = minimum reply time seen (will usually be 14 or 15 if no error)
maxtime = maximum reply time seen (will usually be 14 or 15 if no error)
minerrors = number of times that data read was less than minvalue
maxerrors = number of times that data read was greater than maxvalue
average = the average of all data values read
min = the minimum data value read
max = the maximum data value read

The 'DTC Excitation Voltage' test measures the excitation voltage applied to the bridge of a real DTC scanner. The nominal value for all scanners is 4.5 VDC ± 10%. This equates to a nominal voltage of 4.5VDC = 62259 counts, minimum voltage of 4.05VDC = 59310 counts, and maximum voltage of 4.95VDC = 65208 counts. Therefore, if the 'DTC Excitation Voltage' test is selected, the recommended minvalue/maxvalue range is 59310/65208 with average value of 62304. Also the difference between the min and max should be no more than 25.

If the 'Analog Ground Reference' test is selected the recommended minvalue/maxvalue range is -50 millivolt = 32440 counts to +50 millivolt = 33096 counts. The average value should be near 32767 and the difference between min and max should be no more than 25.

If the 'M0 Logic Zero' test is selected, the recommended minvalue/maxvalue range is 32767/36045. The average value should be near 33496 and the difference between min and max should be no more than 100.

If the 'M1 logic 1' test is selected, the recommended minvalue/maxvalue range is 52429/65535. The average value should be near 56780 and the difference between min and max should be no more than 100.

Example 1: DTC scanner #1 excitation voltage test

XC1 111 6 14 50000 20 1 20 59310 65208
Results: xxxxxxxx "0 14 15 0 0 62304 62298 62310"

Example 2: Analog ground reference test

XC1 111 6 14 50000 20 0 20 32440 33096
Results: xxxxxxxx "0 14 15 0 0 32774 32769 32780"

Example 3: M0 Logic zero test

XC1 111 6 14 50000 20 -1 20 32767 36045
Results: xxxxxxxx "0 14 15 0 0 33496 33492 33502"

Example 4: M1 Logic one test

XC1 111 6 14 50000 20 -2 20 52429 65535
Results: xxxxxxxx "0 14 15 0 0 56780 56755 56788"
 
 

 

 

 
Pressure Systems
34 Research Drive
Hampton, VA 23666
800.678.7226
KPSI Selection Guide Quality Policy
ISO-9001:2000
Receive E-Newsletter
IPC Member Logo WEEE/RoHS Statement
WEEE/RoHS
Statement
 


Home | Environmental Products | Industrial Products | Aerospace Products | Buy Now
KPSI Level and Pressure Selection Guide | Quartzonix Pressure Standard | NetScanner System | System 8400 |
Electronic Pressure Scanners | S8256 System | Miscellaneous Hardware | News Archive
Contact Us | Distributors | Links | Employment | Register | Search