Esterline Pressure Systems Logo header spacer Pressure Systems Site Map Pressure Systems Home
 
 

KPSI Series 750: Specifications

link arrow  return to Series 750 main product page

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
 
Parameter
750
Units
Comments

LEVEL RANGES

Full Scale Level Ranges (1)

10 thru 115
(3 thru 35)

ft H2O
(m H2O)
vented gage reference
Proof Pressure
1.5
x FS
Burst Pressure
2.0
x FS
STATIC PERFORMANCE
Static Accuracy (2)
±0.25
% FSO
BFSL method
Resolution
Infinitesimal

ENVIRONMENTAL

Wetted Materials
316 S; Delrin®; Teflon® PTFE, polyurethane or Viton®
Delrin®, Viton® and Teflon® are registered trademarks of DuPont.
Compensated Temp Range
0 to 50
ºC
Thermal Error (3)

±0.10
 

%FSO/ºC

worst case over compensated temperature range
Operating Temp Range

-20 to 60

0 to 50

ºC

ºC

when using polyurethane cable

when using ETFE cable (5)

Protection Rating

IP 68, NEMA 6P

blank blank

ELECTRICAL

Excitation
9 - 30
VDC
for mA and VDC output
Input Current
20
3.5
mA max
for mA output
for VDC output
Output
4 - 20
0 - 5
mA
VDC
options available (4)
Zero Offset
±0.20
<0.25
mA
VDC
for mA output
for VDC output
Output Impedance

See Loop Resistance diagram page 5 of datasheet
<10


ohm

for mA output

for VDC output
Insulation Resistance
100
mega ohm at 50 VDC
Circuit Protection
Polarity, surge/shorted output

PHYSICAL

Approximate Weight
3.5 (1588)
0.05 (79)
lbs (g)
lbs/ft (g/m)
transducer
cable
Cable
 
Jacket Material


  Pull Strength

  Number of Conductors
 
Conductor Size


Polyurethane (std)
ETFE (6) (opt)

200 (90)

4

22





lbs (kg)



AWG

Tefzel®, Teflon® and Kevlar® are registered trademarks of DuPont.
Cable Seal

Molded polyurethane
Viton® Gland

  for polyurethane cable
for ETFE cable

TEMPERATURE OUTPUT OPTION (Available for 4-20 mA output only)

Temperature Range

0 to 50
-20 to 60

ºC

 
Temperature Output Signal 4 - 20 mA
Temperature Measurement Accuracy ±4 ºC
(1) Intermediate level ranges are available.
(2) Static accuracy includes the combined errors due to nonlinearity, hysteresis and nonrepeatability on a Best Fit Straight Line (BFSL) basis, at 25ºC per ISA S51.1.
(3) Thermal error is the maximum allowable deviation from the Best Fit Straight Line due to a change in temperature, per ISA S51.1.
(4) Custom VDC output levels can be provided up to 2.5 VDC less than the excitation supply voltage.
(5) -20ºC to 50ºC for level ranges <= 100 ft (70m) H2O when using ETFE cable.
(6)

ETFE is a fluoropolymer (Teflon® derivative) material, Tefzel® or equivalent.

Specifications subject to change without notice.
 


 
Pressure Systems
34 Research Drive
Hampton, VA 23666
800.678.7226
KPSI Selection Guide Quality Policy
ISO-9001:2000
Receive E-Newsletter
IPC Member Logo WEEE/RoHS Statement
WEEE/RoHS