Esterline Pressure Systems Logo header spacer Pressure Systems Site Map Pressure Systems Home
 
 

KPSI Series 705: Specifications

link arrow  return to Series 705 main product page

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
   
Parameter
705
Units
Comments

LEVEL RANGES

Full Scale Level Ranges (1)

6 thru 115
(2 thru 35)

ft H2O
(m H2O)
vented gage reference
Proof Pressure
1.5
x FS
Burst Pressure
2.0
x FS

STATIC PERFORMANCE

Static Accuracy (2)
±0.25
% FSO
BFSL method
Resolution
Infinitesimal

ENVIRONMENTAL

Wetted Materials
316 SS or Titanium, Teflon®, polyurethane or Viton®
Viton® and Teflon® are registered trademarks of DuPont.
Compensated Temp Range
0 to 50
ºC
Thermal Error (3, 4)

±0.10
Prorated

%FSO/ºC

for level ranges > 23" (7m) H2O
for level ranges <= 23' (7m) H2O
Operating Temp Range
-20 to 60
ºC
Protection Rating

IP 68, NEMA 6P

ELECTRICAL

Excitation
9 - 30
VDC
for mA and VDC output
Input Current
20
3.5
mA max
for mA output
for VDC output
Output
4 - 20
0 - 5
mA
VDC
options available (5)
Zero Offset
±0.20
<0.1
mA
VDC
for mA output
for VDC output
Output Impedance

See Loop Resistance diagram page 5 of datasheet
<10


ohm

for mA output

for VDC output
Insulation Resistance
100
mega ohm at 50 VDC
Circuit Protection
Polarity, surge/shorted output

PHYSICAL

Approximate Weight
0.5 (227)
0.05 (79)
lbs (g)
lbs/ft (g/m)
transducer
cable
Cable
 
Jacket Material


  Pull Strength

  Number of Conductors
 
Conductor Size


Polyurethane (std)
Tefzel® (opt)

200 (90)

4

22





lbs (kg)



AWG

Tefzel®, Teflon® and Kevlar® are registered trademarks of DuPont.
Cable Seal

Molded polyurethane
Viton® Gland

  for polyurethane cable
for Tefzel® cable

TEMPERATURE OUTPUT OPTION (Not Intrinsic Safety Approved)

Temperature Range

0 to 50
-20 to 60

ºC

available for 4-20 mA output versions only
Output Signal 4 - 20 mA
Temperature Measurement Accuracy ±4 ºC

LIGHTNING PROTECTION (OPTIONAL)

Life Expectancy > 1000 operations    
Peak Clamping Voltage 36 volts    
Response Time < 10 nsecs    
(1) Intermediate level ranges are available.
(2) Static accuracy includes the combined errors due to nonlinearity, hysteresis and nonrepeatability on a Best Fit Straight Line (BFSL) basis, at 25ºC per ISA S51.1.
(3) Thermal error is the maximum allowable deviation from the Best Fit Straight Line due to a change in temperature, per ISA S51.1.
(4) Worst case over compensated temperature range.
(5) Custom VDC output levels can be provided up to 2.5 VDC less than the excitation supply voltage.
   
Specifications subject to change without notice.
 


 
Pressure Systems
34 Research Drive
Hampton, VA 23666
800.678.7226
KPSI Selection Guide Quality Policy
ISO-9001:2000
Receive E-Newsletter
IPC Member Logo WEEE/RoHS Statement
WEEE/RoHS