Parameter
|
Level |
Level
(OSW) |
Barometric Pressure |
Units
|
Comments
|
| 550 |
552 |
551 |
553 |
561 |
570 |
572 |
571 |
573 |
LEVEL RANGES |
| Full Scale Level Ranges (1) |
10 thru 692
(3 thru 210) |
10 thru 50
(3 thru 15) |
N/A |
ft H2O
(m H2O) |
vented
gage reference |
10, 35, 50, 70, 230, 692
(3, 10, 15, 21, 70, 210) |
N/A |
N/A |
ft H2O
(m H2O) |
absolute reference |
| N/A |
8 thru 16
(55 thru 110) |
psia
(kPa) |
absolute reference |
| Proof Pressure |
1.5
|
x FS
|
 |
| Burst Pressure |
2.0
|
x FS
|
 |
MEASUREMENT ACCURACY |
Level (or Pressure) (2, 3) |
±0.05
±0.10 |
N/A |
N/A |
%FS TEB
%FS TEB |
level ranges > 10 ft (3 m) H2O
level ranges = 10 ft (3 m) H2O |
| N/A |
±0.01
±0.1 |
N/A |
ft H2O
% reading |
level ranges <= 10 ft (3 m) H2O
level ranges > 10 ft (3 m) H2O |
| N/A |
N/A |
±0.05 |
%FS TEB |
|
| Temperature |
±1.0 |
±0.2 |
±0.2 |
±1.0 |
±0.2 |
ºC |
 |
DATALOGGING SPECIFICATIONS |
| Modes
of Samping |
Linear, Linear Avg, Event, Logarithmic |
|
software
selectable |
| Units Available |
psi, ft H2O, mm H2O, cm H2O,
m H2O, kPa |
|
user-specified also available by entering slope
and offset |
| Sampling Rate |
3
15 (4) |
rdgs/sec |
typical max |
| Internal Non Volatile Memory |
2 |
4 |
4 |
2 |
4 |
Mbyte |
|
| Maximum Level (or Pressure) Scans |
196,560 |
393,120 |
393,120 |
196,560 |
393,120 |
|
Level (or Pressure) with Time-Stamp |
| Maximum Level (or Pressure) and Tempkerature Scans |
144,144 |
288,288 |
288,288 |
144,144 |
288,288 |
|
Level (or Pressure) and Temperature with Time-Stamp |
| Time-Stamp Accuracy |
±2 |
min/year |
over the compensated temp range |
| Data
Upload Time |
100 |
scans/sec |
@ 19200 baud |
ENVIRONMENTAL |
| Wetted Materials |
316 SS
or Titanium; Viton® |
 |
Viton® is
a registered
trademark of DuPont. |
| Compensated Temp Range |
-10 to 40
|
ºC
|
 |
| Operating Temp Range |
-20 to
50
0 to 50
|
ºC
ºC
|
for all level ranges when using
polyurethane cable or
for level ranges <= 100 ft (70m) H2O
when using ETFE cable
for level ranges > 100 ft (70m) H20
when using ETFE cable |
| Protection Rating |
IP 68, NEMA 6P |
|
|
| Range Storage Temp |
-40 to 80
-20 to 50 |
ºC
ºC |
without
batteries
with batteries |
ELECTRICAL |
| Internal Battery |
2
|
N/A |
2 |
N/A |
2 |
2 |
N/A |
2 |
N/A |
each |
1.5 volt AA, industrial alkaline
recommended |
| Battery Life |
3 |
N/A |
3 |
N/A |
3 |
3 |
N/A |
3 |
N/A |
years |
@ 15
minute sample intervals using industrial alkaline batteries |
| External Excitation |
6-16 |
VDC |
|
| External
Input Current |
8.0
25.0
0.25
|
mA
mA
mA
|
average current during
measurement
15 mS
peak current during page writes quiescent |
| Communication Interface |
RS-485 |
 |
3 volt
p-p differential with selectable baud rates from 1200 to 19200 |
| Protocol |
SDI-12 |
|
ver 1.3
compliant |
PHYSICAL |
| Approximate Weight |
0.70 (318)
0.15 (57)
0.05 (79)
0.005 (8) |
lbs (g)
lbs (g)
lbs/ft (g/m)
lbs/ft (g/m)
|
transducer with batteries
cable assembly (less cable)
cable
suspension wire |
Cable
Jacket Material
Pull Strength
Number of Conductors
Conductor Size |
Polyurethane (standard)
ETFE (opt.) (5)
200 (90)
4
22 |
lbs (kg)
AWG |
Tefzel®, Teflon® and Kevlar® are registered trademarks of DuPont |
| Cable
Seal |
Molded polyurethane
Viton® Gland
|
|
for polyurethane cable
for ETFE cable |
| |
|
| Notes: |
| (1) |
Intermediate level ranges are available. |
|
(2) |
Total Error Band (TEB) includes the combined errors due to nonlinearity, hysteresis, nonrepeatability and thermal effects over the compensated temperature range per ISA S51.1. |
|
(3) |
For absolute reference, additional error up to ±0.2% FS/year may be incurred due to offset drift. |
| (4) |
Leve/temperature averaging must be reduced as required. |
|
(5) |
ETFE is a fluoropolymer ( Teflon® derivative) material, Tefzel® or equivalent. |
| |
|
Specifications subject to change without notice.
|